The CCT189 (body) and CCT191 (head) Low Contrast Phantoms are designed for low contrast imaging buations as defined by the MITA CT IQ Task Group. These phantoms replace the earlier CCT183 phantom.
Both phantoms have eight contrast rods buated at manufacturing to be within 0.05% density or ±0.5 HU of target values. The symmetric bing of the contrast b make these phantoms ideal for conducting forced choice and rating scale experiments for buating reconstruction algorithms.
For further inbation please see the CCT189 and CCT191 MITA CT IQ Low Contrast Phantom Manual.
The CCT189 LC Body Phantom has the following contrast rods:
14HU contrast, 3mm diameter x 40mm length and 15mm diameter x 10mm length 7HU contrast, 5mm diameter x 40mm length and 15mm diameter x 10mm length 5HU contrast, 7mm diameter x 40mm length and 15mm diameter x 10mm length 3HU contrast, 10mm diameter x 40mm length and 15mm diameter x 10mm length
The CCT191 LC Body Phantom has the following contrast rods:
9HU contrast, 3mm diameter x 40mm length and 15mm diameter x 10mm length 7HU contrast, 5mm diameter x 40mm length and 15mm diameter x 10mm length 5HU contrast, 7mm diameter x 40mm length and 15mm diameter x 10mm length 3HU contrast, 10mm diameter x 40mm length and 15mm diameter x 10mm length 深圳為爾康科技有限公司 聯(lián)系人:曾先生 手機 微信號:13718096482 QQ:215620749 電話:0755-28896837 15361860065 網(wǎng)址:www.medicalqc.com