QUART DVT_AP CBCT模體,QUART DVT_AP錐形束CT模體英文介紹: QUART DVT_AP DIN 6868-161 CBCT Image Quality Test Phantom
The QUART DVT_AP phantom is designed to be used as a universal tool for 3D imaging equipment including CT applications. The phantom fully complies with DIN 6868-161 for acceptance tests in CBCT applications.
In conjunction with a specially developed software (QUART DVT_TEC), quick and comprehensive CT IQ tests can be perbed.
Only one exposure required to create 3D data set containing all required beters to buate image quality. Automated buation through unique QUART QA/QC software.Phantom can be applied for field sizes from 4x4cm to large fields-of-view (FOV). A universal holder or customised phantom holders available for easy and reproducible bing.
Technical Specifications SPATIAL RESOLUTION: Line spread function RESOLUTION: Z-Resolution STANDARD TEST OBJECTS: PMMA / Air / PVC MATERIAL EQUIVALENTS: Free Air / Soft tissue / Bone POSTIONING TOOLS: Linear (top side) / Selective markers SIZE: ? 16 cm, height: 15 cm SCATTER RADIATION MODULES: 1x 6 cm / 1x 5 cm Parameters
Nyquist Frequency (NF) Contrast-to-Noise Ratio (CNR) Homogeneity / Image Unibity Z-Resolution Modulation Transfer Function Artefacts, Image Flaws Figure of Merit References EFOMP working group, Quality control in cone-beam computed tomography (CBCT) EFOMP-ESTRO-IAEA protocol (summary report), Physica Medica, Volume 39, July 2017, Pages 67-72.